A dust extraction system has been developed by Lloyd Instruments for its LR50KPlus (50 kN) twin column universal materials testing machine. The system overcomes problems resulting either from the use of dusty samples or dust and small particulates produced during the testing process.  The solution can also be adapted for other twin column material test machines from the Lloyd Instruments range.Exposure to large quantities of dust and small particulates can impair the performance of the instrument. The solution involves the addition of a special filtered blower to the base of the instrument. This creates a positive pressure inside the base unit preventing the dust-laden atmosphere from entering. All cover joints around the base are also sealed.A material testing machine modified in this way has been in use for more than a year and has successfully eliminated any ingress of dust. The filter element needs to be replaced or cleaned every 6 months. The LR50KPlus is ideal for testing applications up to 50 kN (11241 lbf) and is designed equally for routine quality control testing and for performing complex multi-stage tests. The high stiffness frame incorporates a crosshead guidance system to prevent side loading of the test sample. Machine performance can be optimised using NEXYGENPlus material test and data analysis software providing almost unlimited testing capability and results processing features. 

 

Suzanne Sharpe 

Tel. +44 (0)1489 486399 

Fax. +44 (0)1489 885118 

Suzanne.sharpe@ametek.co.uk

A dust extraction system has been developed by Lloyd Instruments for its LR50KPlus (50 kN) twin column universal materials testing machine. The system overcomes problems resulting either from the use of dusty samples or dust and small particulates produced during the testing process.  The solution can also be adapted for other twin column material test machines from the Lloyd Instruments range.

Exposure to large quantities of dust and small particulates can impair the performance of the instrument. The solution involves the addition of a special filtered blower to the base of the instrument. This creates a positive pressure inside the base unit preventing the dust-laden atmosphere from entering. All cover joints around the base are also sealed.

A material testing machine modified in this way has been in use for more than a year and has successfully eliminated any ingress of dust. The filter element needs to be replaced or cleaned every 6 months. 

The LR50KPlus is ideal for testing applications up to 50 kN (11241 lbf) and is designed equally for routine quality control testing and for performing complex multi-stage tests. The high stiffness frame incorporates a crosshead guidance system to prevent side loading of the test sample. Machine performance can be optimised using NEXYGENPlus material test and data analysis software providing almost unlimited testing capability and results processing features. 

Suzanne Sharpe 
Tel. +44 (0)1489 486399 
Fax. +44 (0)1489 885118 
Suzanne.sharpe@ametek.co.uk
PETnology's Resource Guide
comPETence center

The comPETence center provides your organisation with a dynamic, cost effective way to promote your products and services.

Find out more

Current Magazine Cover
Our premium articles
comPETence
magazine

Find our premium articles, interviews, reports and more
in 3 issues in 2026.

Find out more
Current issue